Page 1 - Monitor

Micropaq® MonitorDirections for useModel 406 and Model 408Software version 1.7X

Page 2

6 Chapter 1 General information Welch Allyn Micropaq MonitorIt is possible for the monitor to detect a problem that prevents the monitor from operat

Page 3 - Contents

Directions for use Chapter 1 General information 7Introducing the monitorThe monitor is a patient-worn vital signs monitor for use by adult or pedia

Page 4

8 Chapter 1 General information Welch Allyn Micropaq MonitorUnderstanding the monitor and the FlexNet NetworkThe monitor is intended to operate with

Page 5 - General information

Directions for use Chapter 1 General information 9Monitor featuresControls and connectorsSpO2 ConnectorECG ConnectorVisual Alarm IndicatorLiquid Cry

Page 6

10 Chapter 1 General information Welch Allyn Micropaq MonitorVisual alarm indicatorAudible alarm indicatorBeeps to indicate a patient alarm, and bee

Page 7

Directions for use Chapter 1 General information 11 Scroll Down key and Main Menu keyScrolls down menus on the display.Displays the Main Menu. Selec

Page 8 - General warnings and cautions

12 Chapter 1 General information Welch Allyn Micropaq MonitorDisplayAlthough the Acuity Central Station is the primary location for viewing patient

Page 9

Directions for use Chapter 1 General information 13Main MenuWhen you first press , the Main Menu appears:EXIT Exit the Main Menu (the menu disappea

Page 10

14 Chapter 1 General information Welch Allyn Micropaq MonitorECG SCALE... Change the scale of the ECG waveform. If two waveforms are displayed, both

Page 11 - Introducing the monitor

Directions for use Chapter 1 General information 15Waveform optionsThere are four ECG waveform options as shown:To change the waveform selection dur

Page 12

ii Welch Allyn Micropaq Monitor© 2010 Welch Allyn. All rights are reserved. No one is permitted to reproduce or duplicate, in any form, this manual or

Page 13 - Monitor features

16 Chapter 1 General information Welch Allyn Micropaq MonitorMessages from AcuityThe monitor displays messages sent from Acuity as needed, including

Page 14 - Snapshot key

Directions for use Chapter 1 General information 17Demonstration modeYou can practice using the monitor with the Demo mode of operation, including c

Page 15

18 Chapter 1 General information Welch Allyn Micropaq Monitor5. Press to highlight SERVICE MENU, then press to display the Service Menu.6. Press

Page 16 - Display sleep mode

219MonitoringConnect a new patientConnect to the network1. Insert a battery into the monitor to turn it on. After a few seconds the monitor Power-Up S

Page 17 - Main Menu

20 Chapter 2 Monitoring Welch Allyn Micropaq Monitor3. Press or to highlight the desired Acuity unit, then press .When you press or to highl

Page 18

Directions for use Chapter 2 Monitoring 21• If your patient is in the list, highlight the name and press . Within a few seconds the monitor display

Page 19 - Waveform options

22 Chapter 2 Monitoring Welch Allyn Micropaq Monitor• Even though the monitor contains fully isolated patient-connected circuitry, it has not been s

Page 20 - Operating settings

Directions for use Chapter 2 Monitoring 23Perform 5-Lead ECG monitoring1. Inspect the ECG cable and replace it if it shows any signs of wear, breaka

Page 21 - Demonstration mode

24 Chapter 2 Monitoring Welch Allyn Micropaq Monitor3-Lead ECG application with the 5-Lead ECG cableYou can perform 3-lead ECG monitoring in a simil

Page 22

Directions for use Chapter 2 Monitoring 25Be aware that only ECG lead II is available for display with the monitor’s 3-lead ECG monitoring. No other

Page 23 - Monitoring

iiiContents1 - General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Intended use. . . . . . . . . . . . .

Page 24

26 Chapter 2 Monitoring Welch Allyn Micropaq MonitorPerform SpO2 monitoringWARNING Oxygen saturation measurements using pulse oximetry are highly

Page 25 - Perform ECG monitoring

Directions for use Chapter 2 Monitoring 271. Attach the SpO2 sensor to the patient according to the manufacturer’s directions for use, observing all

Page 26

28 Chapter 2 Monitoring Welch Allyn Micropaq MonitorInstall the carrying pouchAdult carrying pouchThe Adult Carrying Pouch is intended for ambulator

Page 27 - Perform 5-Lead ECG monitoring

Directions for use Chapter 2 Monitoring 29Monitor a patient out of range of AcuityWhile out of range of Acuity, the monitor continues to monitor the

Page 28

30 Chapter 2 Monitoring Welch Allyn Micropaq MonitorStop monitoring a patientIf you want to discontinue monitoring the patient, follow these steps.1

Page 29

Directions for use Chapter 2 Monitoring 31Reconnect a recently monitored patient1. Insert a battery into the monitor to turn on the monitor. Confirm

Page 30 - monitoring

32 Chapter 2 Monitoring Welch Allyn Micropaq MonitorReassign a monitored patient to a new room in the same unitIf a patient is being monitored and y

Page 31

Directions for use Chapter 2 Monitoring 33Transfer a monitored patient to a new room in a different unitIf a patient is being monitored and you want

Page 32 - Install the carrying pouch

34 Chapter 2 Monitoring Welch Allyn Micropaq MonitorReassign the monitor to a new patientIf you want to discontinue monitoring a patient and reconne

Page 33

335Alarms & alertsAbout alarms and alertsAlarms provide a warning about a patient condition (such as a vital sign limit violation).Alerts provide

Page 34 - Stop monitoring a patient

iv Contents Welch Allyn Micropaq Monitor6 - Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51Op

Page 35

36 Chapter 3 Alarms & alerts Welch Allyn Micropaq MonitorRespond to a patient alarm at monitorWhen a patient alarm occurs, the monitor produces

Page 36 - New Room Selection

Directions for use Chapter 3 Alarms & alerts 373. To reset the alarm tone at the monitor and Acuity before the 90 seconds has elapsed, press a

Page 37 - Transfer a Patient

38 Chapter 3 Alarms & alerts Welch Allyn Micropaq Monitor4. To change other limits, scroll to PREVIOUS MENU, press , then select another limit

Page 38 - Select a New Patient

Directions for use Chapter 3 Alarms & alerts 392. If possible, determine what caused the problem and correct it. For a list of possible messages

Page 39 - Alarms & alerts

40 Chapter 3 Alarms & alerts Welch Allyn Micropaq MonitorAlert messages and display informationMessage and Display InformationPossible Cause(s)

Page 40

441Monitor patient at AcuityWhile the Micropaq is connected to the FlexNet network, patient data gathered by the monitor is continuously stored at Acu

Page 41 - Alarm limits can be adjusted

42 Chapter 4 Monitor patient at Acuity Welch Allyn Micropaq Monitor

Page 42

543MaintenanceThis section provides information to help operators of the monitor and the battery charger perform routine maintenance activities such a

Page 43

44 Chapter 5 Maintenance Welch Allyn Micropaq MonitorIf the yellow LED is on continuously, the battery may have reached the end of its useful life.

Page 44

Directions for use Chapter 5 Maintenance 45Inspect the monitor, batteries, battery charger, and accessoriesBefore cleaning, thoroughly inspect the m

Page 45 - Monitor patient at Acuity

11General informationIntended useThe Micropaq® monitor is intended to be used by clinicians for single or multiparameter vital signs monitoring of amb

Page 46

46 Chapter 5 Maintenance Welch Allyn Micropaq MonitorClean the accessoriesClean accessories per manufacturer’s instructions.Approved cleaning soluti

Page 47 - Maintenance

Directions for use Chapter 5 Maintenance 47Recycling monitor componentsWhen the battery, monitor, or battery charger reaches the end of its life, re

Page 48

48 Chapter 5 Maintenance Welch Allyn Micropaq MonitorChange the network nameThis procedure allows you to change the network name assigned to the mon

Page 49

Directions for use Chapter 5 Maintenance 49To change the network name, make sure YES is highlighted, then press to display the following screen:Pr

Page 50 - Clean the accessories

50 Chapter 5 Maintenance Welch Allyn Micropaq Monitor

Page 51 - Recycling monitor components

651ReferenceOperating settingsThe following table lists all of the monitor settings and the default settings.Parameter Set at MonitorSet at Acuity Pre

Page 52 - Change the network name

52 Chapter 6 Reference Welch Allyn Micropaq MonitorSpecificationsMonitor radio specifications (5 GHz)Channel restrictions in the 5-GHz band, by coun

Page 53

Directions for use Chapter 6 Reference 53Monitor radio specifications (2.4 GHz)ECG specificationsThe ECG channel meets all the requirements for Card

Page 54

54 Chapter 6 Reference Welch Allyn Micropaq MonitorCommon Mode RejectionFILTER function OFFFILTER function ON<1 mV p-p RTI for 10V rms, 50/60 Hz

Page 55 - Reference

Directions for use Chapter 6 Reference 55Heart rate and arrhythmia analysis optionMethod for calculating heart rateMonitor Determined by monitor (di

Page 56 - Specifications

2 Chapter 1 General information Welch Allyn Micropaq MonitorSymbolsThe following symbols appear on the monitor or accessories.Warning Warning statem

Page 57 - ECG specifications

56 Chapter 6 Reference Welch Allyn Micropaq MonitorPulse oximetry (SpO2) specifications - MasimoSpO2 complies with EN ISO 9919:2005.Characteristic S

Page 58

Directions for use Chapter 6 Reference 57Pulse oximetry (SpO2) specifications - NellcorSpO2 complies with EN ISO 9919:2005.Characteristic Specificat

Page 59

58 Chapter 6 Reference Welch Allyn Micropaq MonitorPatient alarm and equipment alert specificationsCharacteristic SpecificationVisual Alarm Indicat

Page 60 - ) specifications - Masimo

Directions for use Chapter 6 Reference 59Display specificationsEnvironmental specifications (with battery installed)Transmitter Battery Failure An e

Page 61 - ) specifications - Nellcor

60 Chapter 6 Reference Welch Allyn Micropaq MonitorPhysical specificationsBattery specificationsProtection classifications, all configurations Chara

Page 62

Directions for use Chapter 6 Reference 61Eight-bay battery charger specificationsCharacteristic SpecificationReorder NumberUniversal Battery Charger

Page 63 - Display specifications

62 Chapter 6 Reference Welch Allyn Micropaq Monitor

Page 64 - Battery specifications

763ComplianceGeneralThe 802.11a Wireless PC Card must be installed and used in strict accordance with the manufacturer’s instructions as described in

Page 65

64 Chapter 7 Compliance Welch Allyn Micropaq Monitorsubstitution or attachment of connecting cables and equipment other than specified by Welch Ally

Page 66

Directions for use Chapter 7 Compliance 65Electromagnetic compatibilitySpecial precautions concerning electromagnetic compatibility (EMC) must be ta

Page 67 - Compliance

Directions for use Chapter 1 General information 3Non-ionizing electromagnetic radiation. This device contains an approved RLAN module frequency 240

Page 68 - European Union

66 Chapter 7 Compliance Welch Allyn Micropaq MonitorMonitor Monitor - Guidance and Manufacturer’s Declaration—Electromagnetic EmissionsThe Model 4XX

Page 69 - Electromagnetic compatibility

Directions for use Chapter 7 Compliance 67Monitor - Guidance and Manufacturer’s Declaration—Electromagnetic ImmunityThe Model 4XX Series Monitor is

Page 70

68 Chapter 7 Compliance Welch Allyn Micropaq MonitorMonitor - Guidance and Manufacturer’s Declaration—Electromagnetic ImmunityThe Model 4XX Series M

Page 71

Directions for use Chapter 7 Compliance 69Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the Model 4XX

Page 72

70 Chapter 7 Compliance Welch Allyn Micropaq MonitorBattery charger for the monitor Monitor Battery Charger Only- Guidance and Manufacturer’s Declar

Page 73

Directions for use Chapter 7 Compliance 71Monitor Battery Charger Only- Guidance and Manufacturer’s Declaration—Electromagnetic ImmunityThe Battery

Page 74

72 Chapter 7 Compliance Welch Allyn Micropaq MonitorMonitor Battery Charger Only- Guidance and Manufacturer’s Declaration—Electromagnetic ImmunityTh

Page 75

Directions for use Chapter 7 Compliance 73Recommended Separation Distances between Portable and Mobile RF Communications Equipment and the Battery C

Page 76

74 Chapter 7 Compliance Welch Allyn Micropaq Monitor

Page 77

75Index1-Waveform display, 152-Waveform display, 155-Second waveform display, 15AAccessories, 16Inspection and cleaning, 45Use only recommended, 6, 16

Page 78

4 Chapter 1 General information Welch Allyn Micropaq MonitorBattery charger labels and LEDsGeneral warnings and cautionsFamiliarize yourself with al

Page 79

76 Index Welch Allyn Micropaq MonitorAcuity and settings, 16, 51Operating settings, 16, 51DEFECTIVE SPO2 SENSOR, 40DefibrillationDuring SpO2 monitorin

Page 80

Directions for use Index 77Main MenuEXIT, 13Menu Lock option, 14, 51Restricting access, 14MaintenanceInspection and cleaning, 45Operator maintenance,

Page 81 - Directions for use Index 77

78 Index Welch Allyn Micropaq MonitorReconnect a recently monitored patient, 31Reconnection to network, 8RecyclingBatteries, 47Monitor, 47Red LED, 10R

Page 82

Directions for use Chapter 1 General information 5WARNING Magnetic Resonance (MR) Safety. This monitor is designated Not MR Safe, as defined in th