Micropaq® MonitorDirections for useModel 406 and Model 408Software version 1.7X
6 Chapter 1 General information Welch Allyn Micropaq MonitorIt is possible for the monitor to detect a problem that prevents the monitor from operat
Directions for use Chapter 1 General information 7Introducing the monitorThe monitor is a patient-worn vital signs monitor for use by adult or pedia
8 Chapter 1 General information Welch Allyn Micropaq MonitorUnderstanding the monitor and the FlexNet NetworkThe monitor is intended to operate with
Directions for use Chapter 1 General information 9Monitor featuresControls and connectorsSpO2 ConnectorECG ConnectorVisual Alarm IndicatorLiquid Cry
10 Chapter 1 General information Welch Allyn Micropaq MonitorVisual alarm indicatorAudible alarm indicatorBeeps to indicate a patient alarm, and bee
Directions for use Chapter 1 General information 11 Scroll Down key and Main Menu keyScrolls down menus on the display.Displays the Main Menu. Selec
12 Chapter 1 General information Welch Allyn Micropaq MonitorDisplayAlthough the Acuity Central Station is the primary location for viewing patient
Directions for use Chapter 1 General information 13Main MenuWhen you first press , the Main Menu appears:EXIT Exit the Main Menu (the menu disappea
14 Chapter 1 General information Welch Allyn Micropaq MonitorECG SCALE... Change the scale of the ECG waveform. If two waveforms are displayed, both
Directions for use Chapter 1 General information 15Waveform optionsThere are four ECG waveform options as shown:To change the waveform selection dur
ii Welch Allyn Micropaq Monitor© 2010 Welch Allyn. All rights are reserved. No one is permitted to reproduce or duplicate, in any form, this manual or
16 Chapter 1 General information Welch Allyn Micropaq MonitorMessages from AcuityThe monitor displays messages sent from Acuity as needed, including
Directions for use Chapter 1 General information 17Demonstration modeYou can practice using the monitor with the Demo mode of operation, including c
18 Chapter 1 General information Welch Allyn Micropaq Monitor5. Press to highlight SERVICE MENU, then press to display the Service Menu.6. Press
219MonitoringConnect a new patientConnect to the network1. Insert a battery into the monitor to turn it on. After a few seconds the monitor Power-Up S
20 Chapter 2 Monitoring Welch Allyn Micropaq Monitor3. Press or to highlight the desired Acuity unit, then press .When you press or to highl
Directions for use Chapter 2 Monitoring 21• If your patient is in the list, highlight the name and press . Within a few seconds the monitor display
22 Chapter 2 Monitoring Welch Allyn Micropaq Monitor• Even though the monitor contains fully isolated patient-connected circuitry, it has not been s
Directions for use Chapter 2 Monitoring 23Perform 5-Lead ECG monitoring1. Inspect the ECG cable and replace it if it shows any signs of wear, breaka
24 Chapter 2 Monitoring Welch Allyn Micropaq Monitor3-Lead ECG application with the 5-Lead ECG cableYou can perform 3-lead ECG monitoring in a simil
Directions for use Chapter 2 Monitoring 25Be aware that only ECG lead II is available for display with the monitor’s 3-lead ECG monitoring. No other
iiiContents1 - General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Intended use. . . . . . . . . . . . .
26 Chapter 2 Monitoring Welch Allyn Micropaq MonitorPerform SpO2 monitoringWARNING Oxygen saturation measurements using pulse oximetry are highly
Directions for use Chapter 2 Monitoring 271. Attach the SpO2 sensor to the patient according to the manufacturer’s directions for use, observing all
28 Chapter 2 Monitoring Welch Allyn Micropaq MonitorInstall the carrying pouchAdult carrying pouchThe Adult Carrying Pouch is intended for ambulator
Directions for use Chapter 2 Monitoring 29Monitor a patient out of range of AcuityWhile out of range of Acuity, the monitor continues to monitor the
30 Chapter 2 Monitoring Welch Allyn Micropaq MonitorStop monitoring a patientIf you want to discontinue monitoring the patient, follow these steps.1
Directions for use Chapter 2 Monitoring 31Reconnect a recently monitored patient1. Insert a battery into the monitor to turn on the monitor. Confirm
32 Chapter 2 Monitoring Welch Allyn Micropaq MonitorReassign a monitored patient to a new room in the same unitIf a patient is being monitored and y
Directions for use Chapter 2 Monitoring 33Transfer a monitored patient to a new room in a different unitIf a patient is being monitored and you want
34 Chapter 2 Monitoring Welch Allyn Micropaq MonitorReassign the monitor to a new patientIf you want to discontinue monitoring a patient and reconne
335Alarms & alertsAbout alarms and alertsAlarms provide a warning about a patient condition (such as a vital sign limit violation).Alerts provide
iv Contents Welch Allyn Micropaq Monitor6 - Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51Op
36 Chapter 3 Alarms & alerts Welch Allyn Micropaq MonitorRespond to a patient alarm at monitorWhen a patient alarm occurs, the monitor produces
Directions for use Chapter 3 Alarms & alerts 373. To reset the alarm tone at the monitor and Acuity before the 90 seconds has elapsed, press a
38 Chapter 3 Alarms & alerts Welch Allyn Micropaq Monitor4. To change other limits, scroll to PREVIOUS MENU, press , then select another limit
Directions for use Chapter 3 Alarms & alerts 392. If possible, determine what caused the problem and correct it. For a list of possible messages
40 Chapter 3 Alarms & alerts Welch Allyn Micropaq MonitorAlert messages and display informationMessage and Display InformationPossible Cause(s)
441Monitor patient at AcuityWhile the Micropaq is connected to the FlexNet network, patient data gathered by the monitor is continuously stored at Acu
42 Chapter 4 Monitor patient at Acuity Welch Allyn Micropaq Monitor
543MaintenanceThis section provides information to help operators of the monitor and the battery charger perform routine maintenance activities such a
44 Chapter 5 Maintenance Welch Allyn Micropaq MonitorIf the yellow LED is on continuously, the battery may have reached the end of its useful life.
Directions for use Chapter 5 Maintenance 45Inspect the monitor, batteries, battery charger, and accessoriesBefore cleaning, thoroughly inspect the m
11General informationIntended useThe Micropaq® monitor is intended to be used by clinicians for single or multiparameter vital signs monitoring of amb
46 Chapter 5 Maintenance Welch Allyn Micropaq MonitorClean the accessoriesClean accessories per manufacturer’s instructions.Approved cleaning soluti
Directions for use Chapter 5 Maintenance 47Recycling monitor componentsWhen the battery, monitor, or battery charger reaches the end of its life, re
48 Chapter 5 Maintenance Welch Allyn Micropaq MonitorChange the network nameThis procedure allows you to change the network name assigned to the mon
Directions for use Chapter 5 Maintenance 49To change the network name, make sure YES is highlighted, then press to display the following screen:Pr
50 Chapter 5 Maintenance Welch Allyn Micropaq Monitor
651ReferenceOperating settingsThe following table lists all of the monitor settings and the default settings.Parameter Set at MonitorSet at Acuity Pre
52 Chapter 6 Reference Welch Allyn Micropaq MonitorSpecificationsMonitor radio specifications (5 GHz)Channel restrictions in the 5-GHz band, by coun
Directions for use Chapter 6 Reference 53Monitor radio specifications (2.4 GHz)ECG specificationsThe ECG channel meets all the requirements for Card
54 Chapter 6 Reference Welch Allyn Micropaq MonitorCommon Mode RejectionFILTER function OFFFILTER function ON<1 mV p-p RTI for 10V rms, 50/60 Hz
Directions for use Chapter 6 Reference 55Heart rate and arrhythmia analysis optionMethod for calculating heart rateMonitor Determined by monitor (di
2 Chapter 1 General information Welch Allyn Micropaq MonitorSymbolsThe following symbols appear on the monitor or accessories.Warning Warning statem
56 Chapter 6 Reference Welch Allyn Micropaq MonitorPulse oximetry (SpO2) specifications - MasimoSpO2 complies with EN ISO 9919:2005.Characteristic S
Directions for use Chapter 6 Reference 57Pulse oximetry (SpO2) specifications - NellcorSpO2 complies with EN ISO 9919:2005.Characteristic Specificat
58 Chapter 6 Reference Welch Allyn Micropaq MonitorPatient alarm and equipment alert specificationsCharacteristic SpecificationVisual Alarm Indicat
Directions for use Chapter 6 Reference 59Display specificationsEnvironmental specifications (with battery installed)Transmitter Battery Failure An e
60 Chapter 6 Reference Welch Allyn Micropaq MonitorPhysical specificationsBattery specificationsProtection classifications, all configurations Chara
Directions for use Chapter 6 Reference 61Eight-bay battery charger specificationsCharacteristic SpecificationReorder NumberUniversal Battery Charger
62 Chapter 6 Reference Welch Allyn Micropaq Monitor
763ComplianceGeneralThe 802.11a Wireless PC Card must be installed and used in strict accordance with the manufacturer’s instructions as described in
64 Chapter 7 Compliance Welch Allyn Micropaq Monitorsubstitution or attachment of connecting cables and equipment other than specified by Welch Ally
Directions for use Chapter 7 Compliance 65Electromagnetic compatibilitySpecial precautions concerning electromagnetic compatibility (EMC) must be ta
Directions for use Chapter 1 General information 3Non-ionizing electromagnetic radiation. This device contains an approved RLAN module frequency 240
66 Chapter 7 Compliance Welch Allyn Micropaq MonitorMonitor Monitor - Guidance and Manufacturer’s Declaration—Electromagnetic EmissionsThe Model 4XX
Directions for use Chapter 7 Compliance 67Monitor - Guidance and Manufacturer’s Declaration—Electromagnetic ImmunityThe Model 4XX Series Monitor is
68 Chapter 7 Compliance Welch Allyn Micropaq MonitorMonitor - Guidance and Manufacturer’s Declaration—Electromagnetic ImmunityThe Model 4XX Series M
Directions for use Chapter 7 Compliance 69Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the Model 4XX
70 Chapter 7 Compliance Welch Allyn Micropaq MonitorBattery charger for the monitor Monitor Battery Charger Only- Guidance and Manufacturer’s Declar
Directions for use Chapter 7 Compliance 71Monitor Battery Charger Only- Guidance and Manufacturer’s Declaration—Electromagnetic ImmunityThe Battery
72 Chapter 7 Compliance Welch Allyn Micropaq MonitorMonitor Battery Charger Only- Guidance and Manufacturer’s Declaration—Electromagnetic ImmunityTh
Directions for use Chapter 7 Compliance 73Recommended Separation Distances between Portable and Mobile RF Communications Equipment and the Battery C
74 Chapter 7 Compliance Welch Allyn Micropaq Monitor
75Index1-Waveform display, 152-Waveform display, 155-Second waveform display, 15AAccessories, 16Inspection and cleaning, 45Use only recommended, 6, 16
4 Chapter 1 General information Welch Allyn Micropaq MonitorBattery charger labels and LEDsGeneral warnings and cautionsFamiliarize yourself with al
76 Index Welch Allyn Micropaq MonitorAcuity and settings, 16, 51Operating settings, 16, 51DEFECTIVE SPO2 SENSOR, 40DefibrillationDuring SpO2 monitorin
Directions for use Index 77Main MenuEXIT, 13Menu Lock option, 14, 51Restricting access, 14MaintenanceInspection and cleaning, 45Operator maintenance,
78 Index Welch Allyn Micropaq MonitorReconnect a recently monitored patient, 31Reconnection to network, 8RecyclingBatteries, 47Monitor, 47Red LED, 10R
Directions for use Chapter 1 General information 5WARNING Magnetic Resonance (MR) Safety. This monitor is designated Not MR Safe, as defined in th