Cryosurgery SystemREF 88372Operating Manual
8Care and StorageUpon completion of the surgical procedure, shut off the system at both the console and the tank. A squeeze of the trigger releases an
9SterilizationWARNING: DO NOT AUTOCLAVE THIS INSTRUMENTNOTE: DO ALL CLEANING AND STERILIZATION WITH CLEANING PLUGS AND O-RINGS IN PLACE ON TIPS.COLD S
1088388 Jumbo Exo/Endocervical Cryotip88390 Curved Anorectal Cryotip88392 Condyloma Cryotip88394 Derm Bevel Cryotip88396 Berm Bevel Large Cryotip88398
11WarrantyOne Year Limited Warranty On New InstrumentsInstrumentation purchased new from Welch Allyn, Inc. (Welch Allyn) is warranted to be free from
4341 State Street RoadSkaneateles Falls, NY 13153, U.S.A.Tel.: 1-315-685-4560 or 1-800-535-6663Fax: 1-315-685-3361Printed in the USA 883723 Rev. B
Table of ContentsIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Precautions . . . . . . . . . . . . . . . .
1IntroductionThank you for purchasing the Welch Allyn, Inc. Cryosurgery System for use in your clinic or private practice.Each system is meticulously
2CAUTION: FEDERAL LAW RESTRICTS THE SALE OF THIS DEVICE TO OR ON THE ORDER OF A LICENSED MEDICAL PRACTITIONER OR DESIG-NATED AGENT.WARNING: EXPOSURE T
3getting knocked over. Refill the compressed gas cylinders only by qual-ified producers of compressed gases. Do not allow the storage cylinder tempera
4General InstructionsCheck the gas cylinder to verify that it contains adequate gas. The pressure gauge should read between 600 and 800 PSI for proper
5Directions for Use1. A 20# CO2 tank requries an adapter (supplied). Verify the plastic gas-ket (1) is in place (as shown) prior to fastening the ada
64. Select the proper tip for the cryo procedure and unscrew the protective shield (C) from the end of the cryogun. Hand tighten the cryo tip (D) onto
7training before using this equipment as improper use is potentially haz-ardous to the patient and the user. DO NOT use this device without proper tra