CP 200™ 12-Lead Resting ElectrocardiographDirections for Use

Page 2

6 Chapter 1 Introduction Welch Allyn CP 200 ElectrocardiographAccessoriesTo order accessories, call Welch Allyn. For phone numbers, see page ii.Item

Page 3 - Contents

96 Appendix A Specifications Welch Allyn CP 200 Electrocardiograph*Per AAMI EC11:1991/(R)2001 Diagnostic Electrocardiographic Devices, Section 3.1.2

Page 4

B97EMC Guidance and Manufacturer’s DeclarationsTable 5. Electromagnetic EmissionsThe CP 200 electrocardiograph is intended for use in the electromagn

Page 5 - Introduction

98 Appendix B EMC Guidance and Manufacturer’s Declarations Welch Allyn CP 200 ElectrocardiographTable 6. Electromagnetic ImmunityThe CP 200 electro

Page 6 - Intended Use

Directions for Use Appendix B EMC Guidance and Manufacturer’s Declarations 99Table 7. Electromagnetic ImmunityThe CP 200 electrocardiograph is inte

Page 7 - Indications for Use

100 Appendix B EMC Guidance and Manufacturer’s Declarations Welch Allyn CP 200 ElectrocardiographTable 8. Recommended Separation Distances Between

Page 8

101GlossaryAHA. An abbreviation for American Heart Association. The AHA electrode labeling scheme is most commonly used in the United States. audit tr

Page 9

102 Glossary Welch Allyn CP 200 Electrocardiographlead. (1) An electrocardiograph wire connected to an electrode, which is attached to a patient’s ski

Page 10 - Accessories

103Index3x4, 6x2, 12x1, etc. See leads: arrangements (formats) in Auto Reports3x4 50 mm/s + 3R, 6x2 50 mm/s + 6R. See average cycles5 mm/mV, 10 mm/mV,

Page 11 - Figure 4. Back

104 Index Welch Allyn CP 200 ElectrocardiographDdatabase. See CardioPerfect workstation; scheduled patients list; test directory (saved tests)data. Se

Page 12 - Figure 5. Right Side

Directions for Use Index 105Kkey, network, 36keyboard, 9, 95keys. See function keys; keyboard; softkeysLlanguage setting, 29leads10 lead wires vs. 12

Page 13 - Table 1. Keyboard

Directions for Use Chapter 1 Introduction 7Controls, Indicators, and ConnectorsThis section describes the controls, indicators, and connectors that

Page 14

106 Index Welch Allyn CP 200 ElectrocardiographQQTc method, 38, 46Rreason statements, 43recycling instructions, 87repairs, 94replacement parts and acc

Page 16 - Moving Through the Menus

Reorder Number (multi-language CD): 401151Mat. Number (manual only): 708793, Ver: F4341 State Street Road, PO Box 220, Skaneateles Falls, NY 13153-022

Page 17

8 Chapter 1 Introduction Welch Allyn CP 200 ElectrocardiographFigure 5. Right Side Figure 6. Front Figure 7. Left SideCom port B (USB) SD memory

Page 18

Directions for Use Chapter 1 Introduction 9Table 1. KeyboardKey FunctionA. On/Off See “Powering the Electrocardiograph” on page 24.B. Backspace Del

Page 19 - T2.0A/250V

10 Chapter 1 Introduction Welch Allyn CP 200 ElectrocardiographFigure 8. Softkeys and Function KeysSoftkeysThese softkeys display text or images th

Page 20 - General Warnings

Directions for Use Chapter 1 Introduction 11About the Main MenuThe main menu appears when you press the Menu key . (If you are in spirometry mode wh

Page 21

12 Chapter 1 Introduction Welch Allyn CP 200 ElectrocardiographMoving Through the MenusFigure 10. Standard Menu Figure 11. Parent Menu With Submen

Page 22 - General Cautions

Directions for Use Chapter 1 Introduction 13About the Patient Cable and LeadsThe patient cable processes the patient’s ECG data and transmits it to

Page 23 - Getting Help

14 Chapter 1 Introduction Welch Allyn CP 200 ElectrocardiographSymbolsThe symbols illustrated on the following pages may appear on the electrocardio

Page 24

Directions for Use Chapter 1 Introduction 15Operation SymbolsFor details on the keys, see Figure 1 on page 9.On/standby (off) HelpNavigation arrows

Page 25

ii Welch Allyn CP 200 ElectrocardiographCopyright 2008, Welch Allyn. All rights are reserved. No one is permitted to reproduce or duplicate, in any fo

Page 26 - Connecting the Patient Cable

16 Chapter 1 Introduction Welch Allyn CP 200 ElectrocardiographUsing the Electrocardiograph SafelyBefore using or servicing the electrocardiograph,

Page 27

Directions for Use Chapter 1 Introduction 17Warnings Related to Using the ElectrocardiographWARNING This device captures and presents data reflect

Page 28 - On/Off key

18 Chapter 1 Introduction Welch Allyn CP 200 ElectrocardiographGeneral CautionsThe following caution statements apply to electrocardiograph use in g

Page 29 - Verifying Proper Operation

Directions for Use Chapter 1 Introduction 19Getting HelpYou can get help with the electrocardiograph in a variety of ways beyond this manual.• Press

Page 30

20 Chapter 1 Introduction Welch Allyn CP 200 Electrocardiograph

Page 31 - Reviewing the System Settings

221Setting Up the ElectrocardiographInspecting the Electrocardiograph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Connecting

Page 32 - “System Settings” Menu Tree

22 Chapter 2 Setting Up the Electrocardiograph Welch Allyn CP 200 ElectrocardiographInspecting the Electrocardiograph1. Look for obvious signs of sh

Page 33 - The following screen appears

Directions for Use Chapter 2 Setting Up the Electrocardiograph 23Loading the Thermal Chart Paper 1. Squeeze the latch. Pull the paper door to the le

Page 34

24 Chapter 2 Setting Up the Electrocardiograph Welch Allyn CP 200 ElectrocardiographPowering the ElectrocardiographThe electrocardiograph can run on

Page 35 - 3. Select the desired item:

Directions for Use Chapter 2 Setting Up the Electrocardiograph 25To Turn the Electrocardiograph OnPress . To Turn the Electrocardiograph OffPress a

Page 36

iiiContents1 - Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1About This Manual . . . . . . .

Page 37 - Reviewing the Medication List

26 Chapter 2 Setting Up the Electrocardiograph Welch Allyn CP 200 ElectrocardiographConnecting an External USB Printer (Optional)If desired, you can

Page 38 - Reviewing the History List

327Reviewing the System Settings“System Settings” Menu Tree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28Reviewing the

Page 39

28 Chapter 3 Reviewing the System Settings Welch Allyn CP 200 Electrocardiograph“System Settings” Menu TreeSystem SettingsDevice ConfigurationDevice

Page 40 - Wireless Communication

Directions for Use Chapter 3 Reviewing the System Settings 29Reviewing the Device Configuration Settings1. Press the Menu key .2. Choose System Set

Page 41 - Reviewing the ECG Settings

30 Chapter 3 Reviewing the System Settings Welch Allyn CP 200 ElectrocardiographPressure Unit mmHg, mbar, inHg, kPa. For spirometry only. Determines

Page 42 - “ECG Settings” Menu Tree

Directions for Use Chapter 3 Reviewing the System Settings 31Reviewing the Device Information1. Press the Menu key .2. Choose System Settings >

Page 43 - About Auto ECG Reports

32 Chapter 3 Reviewing the System Settings Welch Allyn CP 200 ElectrocardiographReviewing the Connectivity SettingsSee also “Configuring the Electro

Page 44

Directions for Use Chapter 3 Reviewing the System Settings 33Reviewing the Medication ListThe medication list determines which medications are avail

Page 45

34 Chapter 3 Reviewing the System Settings Welch Allyn CP 200 ElectrocardiographReviewing the History ListThe history list determines which clinical

Page 46

Directions for Use Chapter 3 Reviewing the System Settings 35Transferring a Configuration to Another ElectrocardiographYou can transfer your configu

Page 47

iv Contents Welch Allyn CP 200 ElectrocardiographReviewing the Format Settings for Auto Reports. . . . . . . . . . . . . . . . . . . . . 41Reviewing t

Page 48

36 Chapter 3 Reviewing the System Settings Welch Allyn CP 200 ElectrocardiographConfiguring the Electrocardiograph for Wireless CommunicationBefore

Page 49

437Reviewing the ECG Settings“ECG Settings” Menu Tree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38About Auto ECG

Page 50

38 Chapter 4 Reviewing the ECG Settings Welch Allyn CP 200 Electrocardiograph“ECG Settings” Menu TreeECG SettingsEditAuto Report 1 or 2Auto Report S

Page 51

Directions for Use Chapter 4 Reviewing the ECG Settings 39About Auto ECG ReportsAn Auto ECG is a report of ECG data in one of two user-defined forma

Page 52

40 Chapter 4 Reviewing the ECG Settings Welch Allyn CP 200 ElectrocardiographItem (in Figure 25 on page 39) DescriptionA. Patient data See “Reviewin

Page 53

Directions for Use Chapter 4 Reviewing the ECG Settings 41Reviewing the Format Settings for Auto Reports1. Press the Menu key . 2. Choose ECG Setti

Page 54

42 Chapter 4 Reviewing the ECG Settings Welch Allyn CP 200 ElectrocardiographAverage Cycles If desired, an additional page prints with the report. A

Page 55 - Performing ECG Tests

Directions for Use Chapter 4 Reviewing the ECG Settings 43Reviewing the Interpretation and Copy Settings for Auto Reports1. Press the Menu key . 2.

Page 56

44 Chapter 4 Reviewing the ECG Settings Welch Allyn CP 200 ElectrocardiographReviewing the Patient Data Fields Available for Auto ReportsYou can det

Page 57

Directions for Use Chapter 4 Reviewing the ECG Settings 45Medication Yes or no. If yes, this field is enabled. During data entry, choose up to three

Page 58

11IntroductionAbout This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Product Overview . .

Page 59

46 Chapter 4 Reviewing the ECG Settings Welch Allyn CP 200 ElectrocardiographReviewing the Auto Report Settings1. Press the Menu key . 2. Choose EC

Page 60 - Recording an Auto ECG

Directions for Use Chapter 4 Reviewing the ECG Settings 47Reviewing the Rhythm Report Settings1. Press the Menu key .2. Choose ECG Settings > Rh

Page 61 - Recording a Normal Auto ECG

48 Chapter 4 Reviewing the ECG Settings Welch Allyn CP 200 ElectrocardiographReviewing the Miscellaneous ECG Settings1. Press the Menu key . 2. Cho

Page 62 - Softkeys for adjusting the

Directions for Use Chapter 4 Reviewing the ECG Settings 49Turning the Augmented Pediatric Lead Set On and OffFor a description of the augmented pedi

Page 63 - 8. Press the desired softkey

50 Chapter 4 Reviewing the ECG Settings Welch Allyn CP 200 Electrocardiograph

Page 64

551Performing ECG TestsConnecting the Leads to the Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52Recording an Auto ECG. .

Page 65

52 Chapter 5 Performing ECG Tests Welch Allyn CP 200 ElectrocardiographConnecting the Leads to the Patient1. Help the patient get comfortable. Patie

Page 66 - Recording a Stat Auto ECG

Directions for Use Chapter 5 Performing ECG Tests 53Figure 32. Electrode Placement Locations — Adult and Standard PediatricElectrodesAHA IEC Locati

Page 67 - Recording a Rhythm ECG

54 Chapter 5 Performing ECG Tests Welch Allyn CP 200 ElectrocardiographFigure 33. Electrode Placement Locations — Augmented PediatricElectrodesAHA

Page 68 - Recalling Saved Patient Data

Directions for Use Chapter 5 Performing ECG Tests 554. If the electrocardiograph’s display is blank, press . 5. If prompted to select a lead set typ

Page 69 - Brings up any matching names

2 Chapter 1 Introduction Welch Allyn CP 200 ElectrocardiographAbout This ManualThis manual is written for clinical professionals with a working know

Page 70

56 Chapter 5 Performing ECG Tests Welch Allyn CP 200 ElectrocardiographRecording an Auto ECGAn Auto ECG is a report typically showing a 10-second ac

Page 71 - Today’s scheduled patients

Directions for Use Chapter 5 Performing ECG Tests 57Recording a Normal Auto ECGFor a normal Auto ECG, you enter patient data and do other optional t

Page 72 - Adjusting the ECG Waveforms

58 Chapter 5 Performing ECG Tests Welch Allyn CP 200 ElectrocardiographFigure 38. “Auto ECG Acquisition” Screen4. Verify ECG quality on the screen.

Page 73

Directions for Use Chapter 5 Performing ECG Tests 59Figure 39. “Auto ECG Post-Print” Screen8. Press the desired softkey.Softkey EffectPrint Copy Pr

Page 74

60 Chapter 5 Performing ECG Tests Welch Allyn CP 200 Electrocardiograph Caution The requirements of AAMI EC11, Section 3.2.7.2, Frequency and Impu

Page 75

Directions for Use Chapter 5 Performing ECG Tests 61To Edit or Confirm a Test Directly After Printing1. From the Auto ECG Post-Print screen (Figure

Page 76

62 Chapter 5 Performing ECG Tests Welch Allyn CP 200 ElectrocardiographRecording a Stat Auto ECGA stat Auto ECG is an immediate printout in Auto Rep

Page 77

Directions for Use Chapter 5 Performing ECG Tests 63Recording a Rhythm ECGA Rhythm ECG is a continuous, real-time printout of a rhythm strip with a

Page 78

64 Chapter 5 Performing ECG Tests Welch Allyn CP 200 ElectrocardiographRecalling Saved Patient DataDuring an ECG or spirometry test, instead of ente

Page 79 - Managing Data Security

Directions for Use Chapter 5 Performing ECG Tests 65To Search by Entering a Patient’s Name or ID1. From the Enter New Patient screen (Figure 37 on p

Page 80

Directions for Use Chapter 1 Introduction 3Indications for UseThe electrocardiograph is one of the tools that clinicians use to evaluate, diagnose,

Page 81

66 Chapter 5 Performing ECG Tests Welch Allyn CP 200 ElectrocardiographFigure 44. “Patient Search Results” Screen4. Take the desired action.•Press

Page 82 - Working With the Audit Trail

Directions for Use Chapter 5 Performing ECG Tests 67To Choose a Patient from the Scheduled Patients List1. From the Enter New Patient screen (Figure

Page 83

68 Chapter 5 Performing ECG Tests Welch Allyn CP 200 ElectrocardiographAdjusting the ECG WaveformsThe following softkeys appear on the Auto ECG Acqu

Page 84 - Cleaning the Equipment

669Performing Administrative TasksManaging Saved Tests In the Test Directory . . . . . . . . . . . . . . . . . . . . . . . . . . 70Managing the Schedu

Page 85 - Testing the Equipment

70 Chapter 6 Performing Administrative Tasks Welch Allyn CP 200 ElectrocardiographManaging Saved Tests In the Test DirectoryThe test directory is th

Page 86

Directions for Use Chapter 6 Performing Administrative Tasks 71To Print, Send, or Delete Saved Tests1. Press the Menu key .2. Choose Test Directory

Page 87 - Replacing the Battery

72 Chapter 6 Performing Administrative Tasks Welch Allyn CP 200 ElectrocardiographTo Edit or Confirm Saved Tests1. Select the desired test(s) from t

Page 88

Directions for Use Chapter 6 Performing Administrative Tasks 73Previous Test Saves any changes, and displays the previous test. Appears only when pr

Page 89

74 Chapter 6 Performing Administrative Tasks Welch Allyn CP 200 ElectrocardiographManaging the Scheduled Patients ListThe scheduled patients list id

Page 90 - Replacing the AC Fuses

Directions for Use Chapter 6 Performing Administrative Tasks 75Managing Data SecurityThe CP 200 electrocardiograph incorporates security features th

Page 91 - Discarding the Equipment

4 Chapter 1 Introduction Welch Allyn CP 200 ElectrocardiographPacemaker detectionIf the software detects the possible presence of a pacemaker, it as

Page 92

76 Chapter 6 Performing Administrative Tasks Welch Allyn CP 200 ElectrocardiographWorking With the User List and User LoginThe user list identifies

Page 93 - Troubleshooting

Directions for Use Chapter 6 Performing Administrative Tasks 77To Review or Edit the User List1. Press the Menu key . 2. Choose System Settings >

Page 94 - Problem-Solving Suggestions

78 Chapter 6 Performing Administrative Tasks Welch Allyn CP 200 ElectrocardiographWorking With the Audit TrailAn audit trail, a collection of inform

Page 95

779Maintaining the ElectrocardiographInspecting the Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .80Clean

Page 96

80 Chapter 7 Maintaining the Electrocardiograph Welch Allyn CP 200 ElectrocardiographInspecting the Equipment• Verify that patient leads are fully i

Page 97 - Limited Warranty

Directions for Use Chapter 7 Maintaining the Electrocardiograph 81Testing the EquipmentWhenever the electrocardiograph is serviced or problems are s

Page 98 - Service Policy

82 Chapter 7 Maintaining the Electrocardiograph Welch Allyn CP 200 ElectrocardiographRecharging a Fully Discharged Battery If the electrocardiograph

Page 99 - Specifications

Directions for Use Chapter 7 Maintaining the Electrocardiograph 83Replacing the BatteryIf you have recharged the battery and the electrocardiograph

Page 100

84 Chapter 7 Maintaining the Electrocardiograph Welch Allyn CP 200 Electrocardiograph6. Unplug the two battery connectors. See Figure 53.You may nee

Page 101 - Declarations

Directions for Use Chapter 7 Maintaining the Electrocardiograph 859. Replace the battery door and screws. Turn the electrocardiograph rightside-up.I

Page 102

Directions for Use Chapter 1 Introduction 5OptionsThese options are available both for initial purchases and for upgrades.• SpirometryWith a disposa

Page 103

86 Chapter 7 Maintaining the Electrocardiograph Welch Allyn CP 200 ElectrocardiographReplacing the AC FusesIf the green LED on the keyboard does not

Page 104

Directions for Use Chapter 7 Maintaining the Electrocardiograph 87Storing the EquipmentWhen storing the electrocardiograph, cords, and accessories,

Page 105 - Glossary

88 Chapter 7 Maintaining the Electrocardiograph Welch Allyn CP 200 Electrocardiograph

Page 106

889TroubleshootingProblem-Solving Suggestions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90Limited Warranty . . . . .

Page 107

90 Chapter 8 Troubleshooting Welch Allyn CP 200 ElectrocardiographProblem-Solving SuggestionsThis section includes several tables:• Lead quality pro

Page 108

Directions for Use Chapter 8 Troubleshooting 91 AC interference (even-peaked, regular voltage superimposed on the waveforms). May resemble or coinci

Page 109 - Directions for Use Index 105

92 Chapter 8 Troubleshooting Welch Allyn CP 200 ElectrocardiographTable 4. System Messages (alphabetical order)System Message Problems Actions“A te

Page 110

Directions for Use Chapter 8 Troubleshooting 93Limited WarrantyWelch Allyn, Inc., warrants that the Cardiopulmonary line of electrocardiographs, inc

Page 111

94 Chapter 8 Troubleshooting Welch Allyn CP 200 ElectrocardiographService PolicyAll repairs on products under warranty must be performed or approved

Page 112

A95SpecificationsItem SpecificationDimensionsWeight 11.6 lb (5.3 kg)Keyboard type Elastomer keypad with complete alphanumeric keysSoft-key menu and de