1. General
Applies to Propaq® CS Vital Signs Monitor
Ex: Models 242, 244. Condensed for in situ checkup.
2. Reference Documents
User Guide P/N 810-2689-01 Ver A (online)
Service Line 800-289-2501 Option 3
Service Manual:
http://www.welchallyn.com/documents/Patient
%20Monitoring/Continuous%20Monitoring/Propaq
%20CS/2008_08_12_Propaq_CS_Service_Manual.pdf
3. Tools / Fixtures / Labels
“Smart Arm” for Checking Blood Pressure Cuffs
Pulse Ox Simulator Nellcor SRC-max
EKG Simulator and set of EKG leads - Neurodyne
Green Write-on Label Veriad P/N 171332
Clear Shield Label Veriad P/N US408
4. Basic PM Procedure Note: May be necessary to warn central desk of false alarms.
4.1. Physical Inspection.
Make sure the two small rear circular flanges have not loosened up by temporarily
removing unit from mounting stand.
Check entry of main power cord. When attached, the nearby green “power available”
LED should not flicker if this cord is moved slightly.
Check the external power supply for firm attachment, checking for loose or broken tie-
wraps.
Check the patient leads (EKG, pulse ox, and pressure hose for continuity and integrity).
Note: Hitting any bottom button (actually touch screen icon) will clear alarms.
Central station selection is normally MEDICINE/TRAUMA
4.2. Turn on unit; button switch is on the right side. Unit will normally go thru a set of turn-
on self-diagnostics. Optional: Occasionally necessary to check limit levels of alarms
by following the icons on the touch-screen (hit Setup/Alarms/Limits). When connected
to a network, the upper left of the screen will alternate text “Serial” / “On-network”.
Hit “Network-on/network off”. After a short disconnect interval, unit normally will
reconnect.