Welch Allyn SureTemp 679 Instruction Manual Page 14

  • Download
  • Add to my manuals
  • Print
  • Page
    / 16
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 13
12
the storage channel to clear the display and reset the thermometer to Normal
mode in preparation for another temperature.
TAKING RECTAL TEMPERATURES - RED CAPPED PROBE
1. Exchange the oral probe (blue ejection button) for the rectal probe (red
ejection button) by removing the probe from the storage channel, unplug-
ging the probe connector from the thermometer, and repeating steps 1
through 6 in the SETUP section.
2. Attach a probe cover in the same manner
as in the oral procedure. Make sure the
rEC is displayed.
3. Separate the buttocks with one hand. Us-
ing the other hand, gently insert the probe
only 1.5 cm (5/8 of an inch) inside the
rectal sphincter (less for infants and chil-
dren). The use of lubricant is optional.
WARNING: Incorrect insertion can cause bowel perforation.
Tilt the probe so the tip is touching tissue.
Keep hand separating buttocks in place.
Hold the probe in place during the entire
temperature measurement process.
4. When the final temperature is reached, a
tone will sound and the temperature will
be displayed. Rectal temperature mea-
surements will take about 15 seconds.
5. After the temperature measurement is complete, remove the probe from the
patient, then eject the probe cover by firmly pressing the ejection button on
the probe. (Figure 1).
6. Insert the probe into the storage channel to clear the display, in preparation
for another temperature.
7. Wash hands (note: washing hands greatly reduces the risk of cross-
contamination and NSI).
MAINTENANCE
Other than using a new probe cover for each temperature reading, keeping the
instrument clean, and replacing the batteries, there is no requirement for user
Page view 13
1 2 ... 9 10 11 12 13 14 15 16