Vital Signs Monitor
300 Series
Directions for use
Software version 1.2X
Page view 0
1 2 3 4 5 6 ... 89 90

Summary of Contents

Page 1 - 300 Series

Vital Signs Monitor300 SeriesDirections for useSoftware version 1.2X

Page 2

6 Chapter 1 General Information Welch Allyn Vital Signs Monitor 300 SeriesDisplays, Indicators, Controls, and ConnectionsThis section describes the me

Page 3 - Contents

Directions for Use Chapter 1 General Information 7Status Indicatorsalarms not silencedBattery StatusPulseAmplitude IndicatorºFºFmmHgkPaMMmillimeters o

Page 4

8 Chapter 1 General Information Welch Allyn Vital Signs Monitor 300 SeriesFunction ControlsConnectionsFor information on the connections, refer to the

Page 5 - General Information

29SetupThis chapter describes the set-up procedures for patient monitoring.ConnectionsUse the procedures described below to connect components to the

Page 6 - Table 3. Symbols: Connectors

10 Chapter 2 Setup Welch Allyn Vital Signs Monitor 300 SeriesUse the AC power adapter to fully charge the battery before using the monitor. (This cant

Page 7 - Product Overview

Directions for Use Chapter 2 Setup 11Connecting the NIBP Cuff HoseAttach the hose to the monitor and the cuff as follows, referring to the illustratio

Page 8 - Warnings and Cautions

12 Chapter 2 Setup Welch Allyn Vital Signs Monitor 300 SeriesConnecting the Temperature Probe CableFollow these steps to connect the temperature probe

Page 9 - General Cautions

Directions for Use Chapter 2 Setup 13Connecting and Disconnecting the SpO2Sensor CableTo connect the SpO2 sensor cable:1. Locate the SpO2sensor cable

Page 10

14 Chapter 2 Setup Welch Allyn Vital Signs Monitor 300 SeriesTo disconnect the SpO2cable, refer to the instructions shown in the illustration below.Fo

Page 11 - Status Indicators

Directions for Use Chapter 2 Setup 15Power On, Power-on Self-Test, and Power OffWhen the battery is charged, press to turn on the monitor.The monitor

Page 12 - 30V , 1A Max

ii Welch Allyn Vital Signs Monitor 300 Series© 2014 Welch Allyn. All rights are reserved. To support the intended use of the product described in this

Page 13 - Connections

16 Chapter 2 Setup Welch Allyn Vital Signs Monitor 300 SeriesConfiguring Operating ParametersYou can change several monitor operating parameters. When

Page 14 - Adapter Port

Directions for Use Chapter 2 Setup 17Use the settings menu to select and set the following parameters:To change a settings parameter:1. Select the par

Page 15 - Connecting the NIBP Cuff Hose

18 Chapter 2 Setup Welch Allyn Vital Signs Monitor 300 SeriesConfiguration MenuThe configuration menu is accessed by pressing and keeping it depressed

Page 16 - TemperatureProbe

Directions for Use Chapter 2 Setup 19Use the configuration menu to select and set the following parameters:To change a configuration parameter:1. Sele

Page 17 - Sensor Cable

20 Chapter 2 Setup Welch Allyn Vital Signs Monitor 300 SeriesChanging the Time and DateFollow these steps to change the time and date settings of the

Page 18

Directions for Use Chapter 2 Setup 2111. To save the displayed time and date settings, either do nothing for 10 seconds orpress any button other than

Page 19 - ****************

22 Chapter 2 Setup Welch Allyn Vital Signs Monitor 300 SeriesChanging the Patient TypeThe age range for each patient type is defined as follows:Defaul

Page 20 - How to Use the Menu System

Directions for Use Chapter 2 Setup 23MAP Measurement Enable and DisableDefault setting: MAP ENABLED for neonate; MAP DISABLED for adult and pediatric.

Page 21

24 Chapter 2 Setup Welch Allyn Vital Signs Monitor 300 SeriesChanging the NIBP Measurement UnitsDefault setting: mmHg.To change the NIBP measurement u

Page 22 - MAP 90 kPa

Directions for Use Chapter 2 Setup 25Changing Temperature Type and Measurement UnitsDefault setting: F (Fahrenheit predictive).To change the temperatu

Page 23

iiiContents1 - General Information ...1About This Manual ...1Intended

Page 24 - Changing the Time and Date

26 Chapter 2 Setup Welch Allyn Vital Signs Monitor 300 SeriesChanging the Volume of the Pulse ToneDefault setting: 03.The pulse tone can be set from l

Page 25 - ERASE DATA?

Directions for Use Chapter 2 Setup 27Selecting Stream or Batch PrintingDefault setting: BATCH.For monitors configured with the optional thermal printe

Page 26 - Changing the Patient Type

28 Chapter 2 Setup Welch Allyn Vital Signs Monitor 300 Series

Page 27 - MAP ENABLED

329Patient MonitoringMonitoring Blood PressureWarnings — NIBPWARNING To ensure safe and accurate NIBP measurements, use only cuffs andhoses approved b

Page 28 - BP Units: kPa

30 Chapter 3 Patient Monitoring Welch Allyn Vital Signs Monitor 300 SeriesNIBP PreparationBefore you start any NIBP measurement, always follow the ste

Page 29 - 20:30:16

Directions for Use Chapter 3 Patient Monitoring 31If Action 1 and Action 2 lead to Condition 1 and Condition 2, do the following to restorethe monitor

Page 30 - VOLUME 05

32 Chapter 3 Patient Monitoring Welch Allyn Vital Signs Monitor 300 SeriesManual NIBP MeasurementFollow these steps to take a single NIBP measurement.

Page 31 - PRINT: BATCH

Directions for Use Chapter 3 Patient Monitoring 33• If MAP is enabled, MAP results are displayed in the message window.The measurement display persist

Page 32

34 Chapter 3 Patient Monitoring Welch Allyn Vital Signs Monitor 300 Series4. To select the currently displayed interval, press any button other than ,

Page 33 - Patient Monitoring

Directions for Use Chapter 3 Patient Monitoring 35The automatic NIBP cycles continue until one of the following occurs:• The monitor reaches the 5-min

Page 34 - NIBP Preparation

iv Contents Welch Allyn Vital Signs Monitor 300 Series7 - Reference ...69Battery Operation ...

Page 35 - Positioning the Cuff

36 Chapter 3 Patient Monitoring Welch Allyn Vital Signs Monitor 300 SeriesHow Changing the Patient Type Affects MAP DefaultsWhen you cycle power to th

Page 36 - Manual NIBP Measurement

Directions for Use Chapter 3 Patient Monitoring 37Monitoring SpO2Warnings and Cautions — SpO2WARNING Always follow the manufacturer’s instructions for

Page 37 - Automatic NIBP Measurement

38 Chapter 3 Patient Monitoring Welch Allyn Vital Signs Monitor 300 SeriesSpO2Monitoring Procedure1. Verify that the SpO2sensor cable is connected to

Page 38 - 20:05:42

Directions for Use Chapter 3 Patient Monitoring 39Within a few seconds, the pulse amplitude indicator reflects the rate and strength ofthe pulse.Withi

Page 39 - 10.7 9 8

40 Chapter 3 Patient Monitoring Welch Allyn Vital Signs Monitor 300 SeriesMonitoring TemperatureWarnings and Cautions — TemperatureWARNING To ensure p

Page 40 - Monitoring Pulse Rate

Directions for Use Chapter 3 Patient Monitoring 41Setting the Temperature Measurement TypeThe monitor, if configured with the temperature option, can

Page 41 - Monitoring SpO

42 Chapter 3 Patient Monitoring Welch Allyn Vital Signs Monitor 300 SeriesLoading a Probe Cover1. Holding the probe handle with your thumb and two fin

Page 42 - Monitoring Procedure

Directions for Use Chapter 3 Patient Monitoring 43Oral PredictiveWhen used correctly, the monitor produces an accurate oral temperature measurement in

Page 43 - 17.1 9 9

44 Chapter 3 Patient Monitoring Welch Allyn Vital Signs Monitor 300 Series5. Hold the probe in place to assure continuous contact with the oral tissue

Page 44 - Monitoring Temperature

Directions for Use Chapter 3 Patient Monitoring 45Axillary PredictiveWhen used correctly, the monitor produces an accurate axillary temperaturemeasure

Page 45

11General InformationAbout This ManualThis manual contains information about the Welch Allyn®Vital Signs Monitor 300 Seriesmonitor. The series include

Page 46 - Ejecting a Used Probe Cover

46 Chapter 3 Patient Monitoring Welch Allyn Vital Signs Monitor 300 Series5. Place the probe tip as high as possible in the axilla, and then bring the

Page 47 - Oral Predictive

Directions for Use Chapter 3 Patient Monitoring 47Rectal PredictiveWhen used correctly, the monitor produces an accurate rectal temperature measuremen

Page 48

48 Chapter 3 Patient Monitoring Welch Allyn Vital Signs Monitor 300 SeriesMonitored Temperature MeasurementThe procedures for monitored and predictive

Page 49 - Axillary Predictive

Directions for Use Chapter 3 Patient Monitoring 49Thermometer and Probe Cleaning Procedure1. Wipe the thermometer regularly with a cloth dampened with

Page 50

50 Chapter 3 Patient Monitoring Welch Allyn Vital Signs Monitor 300 Series

Page 51 - Rectal Predictive

451Alarms and AlertsResponding to a Patient AlarmA patient alarm occurs when a vital-sign measurement falls outside of programmed limits.During a pati

Page 52 - 20:27:50

52 Chapter 4 Alarms and Alerts Welch Allyn Vital Signs Monitor 300 SeriesResponding to an Equipment AlertRecoverable Temperature, NIBP, or SpO2Alert—N

Page 53

Directions for Use Chapter 4 Alarms and Alerts 53Nonrecoverable AlertsWhen the monitor detects a nonrecoverable equipment problem, it does the followi

Page 54

54 Chapter 4 Alarms and Alerts Welch Allyn Vital Signs Monitor 300 SeriesAlarm IndicatorsThe monitor alarm indicators are as follows:Setting AlarmsDur

Page 55 - Alarms and Alerts

Directions for Use Chapter 4 Alarms and Alerts 55Set alarms for systolic and diastolic blood pressure, pulse rate, and SpO2as follows:1. Press .• All

Page 56 - Alert—Escalated

2 Chapter 1 General Information Welch Allyn Vital Signs Monitor 300 SeriesSymbolsThe symbols illustrated on the following pages appear on the monitor

Page 57 - Nonrecoverable Alerts

56 Chapter 4 Alarms and Alerts Welch Allyn Vital Signs Monitor 300 Series6. Change or accept the MAP high alarm limit as described above (from step 2)

Page 58 - Setting Alarms

Directions for Use Chapter 4 Alarms and Alerts 57Nurse CallThe monitor can be connected to a Nurse Call system through a customized cable thatconnects

Page 59

58 Chapter 4 Alarms and Alerts Welch Allyn Vital Signs Monitor 300 SeriesError CodesErrorCodeDescriptionBlood PressureC01 NIBP cycle cancelled by oper

Page 60

559Reviewing Patient DataYou can review stored patient data either by viewing it on the monitor or by printing it.Displaying Stored Patient Data1. Pre

Page 61 - Nurse Call

60 Chapter 5 Reviewing Patient Data Welch Allyn Vital Signs Monitor 300 SeriesTo Start and Stop Printing• If the monitor is not printing, press to sta

Page 62 - Error Codes

Directions for Use Chapter 5 Reviewing Patient Data 61• is pressed again; the monitor prints an error log (if any errors occurred) and afooter, and th

Page 63 - Reviewing Patient Data

62 Chapter 5 Reviewing Patient Data Welch Allyn Vital Signs Monitor 300 SeriesTable HeadingThe Table Heading contains some or all of the following col

Page 64 - Stream Printing

Directions for Use Chapter 5 Reviewing Patient Data 63Range and Alarm FlagsFlag characters ( and ) indicate a data value that falls outside of a def

Page 65 - Printer Output

64 Chapter 5 Reviewing Patient Data Welch Allyn Vital Signs Monitor 300 SeriesNIBPNIBP notation can include error information:Error LegendIf any error

Page 66 -  NOR/ORL

Directions for Use Chapter 5 Reviewing Patient Data 65Erasing Patient DataAll patient vital-sign data is erased when the monitor is powered off or whe

Page 67

Directions for Use Chapter 1 General Information 3The functions of the monitor front panel controls illustrated here are described in detailelsewhere

Page 68 - Error Legend

66 Chapter 5 Reviewing Patient Data Welch Allyn Vital Signs Monitor 300 SeriesReplacing the Printer Paper SupplyTo replace the printer paper supply, f

Page 69 - Erasing Patient Data

667Operator MaintenanceThis chapter describes the operator maintenance procedures.CleaningWipe the monitor with a cloth slightly dampened with warm wa

Page 70

68 Chapter 6 Operator Maintenance Welch Allyn Vital Signs Monitor 300 SeriesRecycling Monitor ComponentsWithin the EUDo not dispose of this product as

Page 71 - Operator Maintenance

769ReferenceBattery OperationBattery Low WarningWhen the battery charge is low, a warning tone sounds and the Low Battery indicatorflashes continuousl

Page 72 - Recycling Monitor Components

70 Chapter 7 Reference Welch Allyn Vital Signs Monitor 300 SeriesBattery failure causes the immediate suspension of all monitoring and print functions

Page 73 - Reference

Directions for Use Chapter 7 Reference 71Monitor SpecificationsThe monitor is an FDA Class II non-critical device and, according to Council Directive

Page 74 - Battery Replacement

72 Chapter 7 Reference Welch Allyn Vital Signs Monitor 300 SeriesCharacteristic Specification NotesTemperatureTemperatureMeasurement Range80 Fto110F

Page 75 - Monitor Specifications

Directions for Use Chapter 7 Reference 73PhysicalCharacteristic SpecificationDimensionsheight 6.6 inches 16.8 cmwidth 10.0 inches 25.4 cmdepth 6.0 inc

Page 76

74 Chapter 7 Reference Welch Allyn Vital Signs Monitor 300 SeriesElectricalPower RequirementsThe medical-rated isolation AC power adapter connected to

Page 77 - Physical

Directions for Use Chapter 7 Reference 75EnvironmentalEMC ComplianceThe monitor complies with all applicable and required standards for electromagneti

Page 78 - Electrical

4 Chapter 1 General Information Welch Allyn Vital Signs Monitor 300 Seriesmedical equipment). All such configurations must comply with system standard

Page 79 - Environmental

76 Chapter 7 Reference Welch Allyn Vital Signs Monitor 300 SeriesGuidance and manufacturer’s declaration—electromagnetic immunityThe Vital Signs Monit

Page 80

Directions for Use Chapter 7 Reference 77Guidance and manufacturer’s declaration—electromagnetic immunityThe VSM 300 Series monitor is intended for us

Page 81

78 Chapter 7 Reference Welch Allyn Vital Signs Monitor 300 SeriesOther Environmental SpecificationsRecommended separation distances between portable a

Page 82

Directions for Use Chapter 7 Reference 79Nurse CallSwitch current 1A maximumSwitch voltage 30V ac/dc maximumIsolation 1500 VrmsAlarm relay Energized d

Page 83

80 Chapter 7 Reference Welch Allyn Vital Signs Monitor 300 SeriesFactory Default SettingsFunction ValueBlood pressure measurement units mmHgPatient ty

Page 84 - Factory Default Settings

81Limited WarrantyThis product is sold by Welch Allyn under the warranties set forth in the followingparagraphs. These warranties are extended only to

Page 85 - Limited Warranty

82 Limited Warranty Welch Allyn Vital Signs Monitor 300 Series

Page 86

83IndexAAC powerconnecting, 9adultsee also patient typedefined, 22icon, 3alarmhigh and low, 55MAP, 34, 56patient, 51range limits, 56responding to, 51s

Page 87

84 Index Welch Allyn Vital Signs Monitor 300 Seriesdescription, 7intended use, 1intravenous infusionwarning, 29KkPaunits, selecting, 24Mmaintenance, 6

Page 88

Directions for Use Index 85Vvital signsalarms, 51blood pressure, 29pulse rate, 36SpO2, 37temperature, 40Wwarningsgeneral, 4NIBP, 29SpO2, 37temperature

Page 89

Directions for Use Chapter 1 General Information 5General CautionsWARNING Do not connect more than one patient to a monitor.WARNING If the monitor is

Page 90

86 Index Welch Allyn Vital Signs Monitor 300 Series