Welch-allyn Spot Vital Signs LXI - Quick Reference Guide User Manual

Browse online or download User Manual for Equipment Welch-allyn Spot Vital Signs LXI - Quick Reference Guide. Welch Allyn Spot Vital Signs LXI - Quick Reference Guide User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 2
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
Blood Pressure Operation
Press the BP Start/Stop button to start a blood pressure
measurement; press it again to stop a current reading.
SureTemp Plus Temperature Operation
1. Remove probe from well and apply a new probe cover. Use the
Navigation buttons to change the temperature mode then
press the Select button.
2. Place probe in contact with patient tissue as directed in the
Directions for Use manual. When done, a beep sounds. The temperature appears
on the display.
3. Eject probe cover. Return probe to well before starting another measurement.
Braun ThermoScan
®
PRO 4000 Temperature Operation
1. Remove probe from holder and apply a new probe cover; when ready, the device
will beep.
2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
3. Press the Activation button. When done, a long beep sounds. The probe will
display the temperature.
4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will
display the temperature.
SpO
2
Operation
Apply sensor to patient. Verify there is a signal on the pulse signal bar graph. The
SpO
2
measurement and pulse rate will update continuously for 10 minutes.
Blood Pressure Operation
Press the BP Start/Stop button to start a blood pressure
measurement; press it again to stop a current reading.
SureTemp Plus Temperature Operation
1. Remove probe from well and apply a new probe cover. Use the
Navigation buttons to change the temperature mode then
press the Select button.
2. Place probe in contact with patient tissue as directed in the
Directions for Use manual. When done, a beep sounds. The temperature appears
on the display.
3. Eject probe cover. Return probe to well before starting another measurement.
Braun ThermoScan
®
PRO 4000
Temperature Operation
1. Remove probe from holder and apply a new probe cover; when ready, the device
will beep.
2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
3. Press the Activation button. When done, a long beep sounds. The probe will
display the temperature.
4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will
display the temperature.
SpO
2
Operation
Apply sensor to patient. Verify there is a signal on the pulse signal bar graph. The
SpO
2
measurement and pulse rate will update continuously for 10 minutes.
Blood Pressure Operation
Press the BP Start/Stop button to start a blood pressure
measurement; press it again to stop a current reading.
SureTemp Plus Temperature Operation
1. Remove probe from well and apply a new probe cover. Use the
Navigation buttons to change the temperature mode then
press the Select button.
2. Place probe in contact with patient tissue as directed in the
Directions for Use manual. When done, a beep sounds. The temperature appears
on the display.
3. Eject probe cover. Return probe to well before starting another measurement.
Braun ThermoScan
®
PRO 4000 Temperature Operation
1. Remove probe from holder and apply a new probe cover; when ready, the device
will beep.
2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
3. Press the Activation button. When done, a long beep sounds. The probe will
display the temperature.
4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will
display the temperature.
SpO
2
Operation
Apply sensor to patient. Verify there is a signal on the pulse signal bar graph. The
SpO
2
measurement and pulse rate will update continuously for 10 minutes.
Blood Pressure Operation
Press the BP Start/Stop button to start a blood pressure
measurement; press it again to stop a current reading.
SureTemp Plus Temperature Operation
1. Remove probe from well and apply a new probe cover. Use the
Navigation buttons to change the temperature mode then
press the Select button.
2. Place probe in contact with patient tissue as directed in the
Directions for Use manual. When done, a beep sounds. The temperature appears
on the display.
3. Eject probe cover. Return probe to well before starting another measurement.
Braun ThermoScan
®
PRO 4000 Temperature Operation
1. Remove probe from holder and apply a new probe cover; when ready, the device
will beep.
2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
3. Press the Activation button. When done, a long beep sounds. The probe will
display the temperature.
4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will
display the temperature.
SpO
2
Operation
Apply sensor to patient. Verify there is a signal on the pulse signal bar graph. The
SpO
2
measurement and pulse rate will update continuously for 10 minutes.
Spot Vital Signs LXi
Operating Instructions
On/Off Button
Controls power to the device.
Select Button
Accepts the current option.
Spot Vital Signs LXi
Operating Instructions
On/Off Button
Controls power to the device.
Select Button
Accepts the current option.
Spot Vital Signs LXi
Operating Instructions
On/Off Button
Controls power to the device.
Select Button
Accepts the current option.
Spot Vital Signs LXi
Operating Instructions
On/Off Button
Controls power to the device.
Select Button
Accepts the current option.
Blood Pressure Operation
Press the BP Start/Stop button to start a blood pressure
measurement; press it again to stop a current reading.
SureTemp Plus Temperature Operation
1. Remove probe from well and apply a new probe cover. Use the
Navigation buttons to change the temperature mode then press
the Select button.
2. Place probe in contact with patient tissue as directed in the
Directions for Use manual. When done, a beep sounds. The temperature appears
on the display.
3. Eject probe cover. Return probe to well before starting another measurement.
Braun ThermoScan
®
PRO 4000 Temperature Operation
1. Remove probe from holder and apply a new probe cover; when ready, the device
will beep.
2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
3. Press the Activation button. When done, a long beep sounds. The probe will
display the temperature.
4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will
display the temperature.
SpO
2
Operation
Apply sensor to patient. Verify there is a signal on the pulse signal bar graph. The
SpO
2
measurement and pulse rate will update continuously for 10 minutes.
Blood Pressure Operation
Press the BP Start/Stop button to start a blood pressure
measurement; press it again to stop a current reading.
SureTemp Plus Temperature Operation
1. Remove probe from well and apply a new probe cover. Use the
Navigation buttons to change the temperature mode then press
the Select button.
2. Place probe in contact with patient tissue as directed in the
Directions for Use manual. When done, a beep sounds. The temperature appears
on the display.
3. Eject probe cover. Return probe to well before starting another measurement.
Braun ThermoScan
®
PRO 4000 Temperature Operation
1. Remove probe from holder and apply a new probe cover; when ready, the device
will beep.
2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
3. Press the Activation button. When done, a long beep sounds. The probe will
display the temperature.
4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will
display the temperature.
SpO
2
Operation
5. Apply sensor to patient. Verify there is a signal on the pulse signal bar graph. The
SpO
2
measurement and pulse rate will update continuously for 10 minutes.
Blood Pressure Operation
Press the BP Start/Stop button to start a blood pressure
measurement; press it again to stop a current reading.
SureTemp Plus Temperature Operation
1. Remove probe from well and apply a new probe cover. Use the
Navigation buttons to change the temperature mode then
press the Select button.
2. Place probe in contact with patient tissue as directed in the
Directions for Use manual. When done, a beep sounds. The temperature appears
on the display.
3. Eject probe cover. Return probe to well before starting another measurement.
Braun ThermoScan
®
PRO 4000 Temperature Operation
1. Remove probe from holder and apply a new probe cover; when ready, the device
will beep.
2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
3. Press the Activation button. When done, a long beep sounds. The probe will
display the temperature.
4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will
display the temperature.
SpO
2
Operation
Apply sensor to patient. Verify there is a signal on the pulse signal bar graph. The
SpO
2
measurement and pulse rate will update continuously for 10 minutes.
Blood Pressure Operation
Press the BP Start/Stop button to start a blood pressure
measurement; press it again to stop a current reading.
SureTemp Plus Temperature Operation
1. Remove probe from well and apply a new probe cover. Use the
Navigation buttons to change the temperature mode then
press the Select button.
2. Place probe in contact with patient tissue as directed in the
Directions for Use manual. When done, a beep sounds. The temperature appears
on the display.
3. Eject probe cover. Return probe to well before starting another measurement.
Braun ThermoScan
®
PRO 4000 Temperature Operation
1. Remove probe from holder and apply a new probe cover; when ready, the device
will beep.
2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
3. Press the Activation button. When done, a long beep sounds. The probe will
display the temperature.
4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will
display the temperature.
SpO
2
Operation
Apply sensor to patient. Verify there is a signal on the pulse signal bar graph. The
SpO
2
measurement and pulse rate will update continuously for 10 minutes.
Spot Vital Signs LXi
Operating Instructions
On/Off Button
Controls power to the device.
Select Button
Accepts the current option.
Spot Vital Signs LXi
Operating Instructions
On/Off Button
Controls power to the device.
Select Button
Accepts the current option.
Spot Vital Signs LXi
Operating Instructions
On/Off Button
Controls power to the device.
Select Button
Accepts the current option.
Spot Vital Signs LXi
Operating Instructions
On/Off Button
Controls power to the device.
Select Button
Accepts the current option.
Page view 0
1 2

Summary of Contents

Page 1

Blood Pressure OperationPress the BP Start/Stop button to start a blood pressure measurement; press it again to stop a current reading.SureTemp Plus T

Page 2

Save/Send Operation1. Use the Navigation buttons to move the arrow to “Save/Next Reading” or “Send/Next Reading” in the Navigation window. 2. Press t

Comments to this Manuals

No comments