ProBP 3400 tab
Calibrate and verify the device1.Click within the device row, then click Perform All to activate the test sequence for the module.2.Enter information
Configure tabThe configure tab enables all users the option to change the location ID and the asset tag number.Additionally, users may configure the P
ContentsProBP 3400 tab ...
ProBP 3400 tabThe ProBP 3400 tab is the single access point to all device information and actions that you may perform on a ProBP3400 (except product
Managed devicesWelch Allyn Service Tool provides features that can help you manage and maintain your inventory of supportedWelch Allyn products. The W
ProBP3400 error codesThe following tables list the error messages associated with the Welch Allyn ProBP 3400.Error code Related System Additional deta
Error condition Definition / Description Required actionSet date and time. Either date or time not set at all ordata has not been set.Set date and tim
Error condition Definition / Description Required actionDevice is operating in battery mode. AC power cord has beendisconnected.Click the OK button to
ProBP 3400 calibration and verification setupTest equipment part numbersThe equipment detailed in the table below are used for ProBP 3400 calibration
1.Connect a USB cable to the service test box; connect the opposite end to a PC.2.Connect a mini B USB cable to the ProBP 3400 and the opposite end to