1.
Connect a USB cable to the service test box; connect the opposite end to a PC.
2.
Connect a mini B USB cable to the ProBP 3400 and the opposite end to the service test box.
3.
Connect the blood pressure hose to the NIBP module and the test volume.
Connect a piece of tubing to the squeeze bulb with bleeding screw; connect the opposite end to the square
fitting connector.
Connect another piece of tubing to the square fitting connector; connect the opposite end to the calibrated
pressure meter.
Connect another piece of tubing to the square fitting connector; connect the opposite end to the test volume.
Connect the last piece of tubing to the square fitting connector; connect the opposite end to the "Y" fitting on
the two-tube blood pressure connector.
OpenTopic | ProBP 3400 calibration and verification setup | 9